Figure 21: Disinfection cell with mounted strainer
- Attach the strainer with the ClO
2
-tablet into the housing of the disinfection cell. To
do that press the strainer into the fit inside of the housing.
- Close the disinfection cell by turning the cell case onto the cell-housing
counterclockwise
6.8 Replacement of reverse osmosis module
- Turn off the system by unplugging the power adapter out of the electrical outlet.
- Disconnect the system from feed water.
- Open the system.
- Remove reverse osmosis housing from its holder.
- Remove tubes from the reverse osmosis housing (caution: water leakage).
- Open the housing and remove the reverse osmosis module.
- Unpack the new reverse osmosis module and put it into the housing.
- Close the housing.
- Reconnect the tubes onto the reverse osmosis housing.
- Plug the reverse osmosis housing back into its holders.
- Connect the system to feed water.
- Turn on the system by plugging the power adapter into an electrical outlet.
- Flush the reverse osmosis module by start the tool “CHANGE ProPak.” (see chapter
5.2.1).
- Check impermeability on all connections
- Close the front door, the system is ready for operation.