62
Chapter 6 Lamp Replacement
VISERA XENON LIGHT SOURCE CLV-S40
6. Disengage the clamp on heat sink (a) and pull out the lamp (see Figure 6.4).
7. Using a clean lint-free cloth, wipe off any residual heat compound from the
heat sink.
8. Remove the new lamp from its container.
9. With a finger, apply the supplied heat compound (for better heat conduction)
evenly and thickly on the area indicated by hatched lines on the “–” side of
the lamp in Figure 6.5. Then insert the lamp’s “–” side into heat sink (a) as
far as it will go.
Apply a sufficient amount of heat compound. Otherwise, the lamp may fail to
light after several hours.
Figure 6.5
10. Apply heat compound to the hatched area on the “+” surface of the lamp in
the same manner. Then, fit heat sink (b) over the lamp’s “+” side. Tighten
the three bolts firmly using the hexagon wrench.
Align the guide pin on the heat sink (b) with the guide pin hole, and tighten
the bolts securely.
• Do not smear the lamp’s glass surface. Wipe off excess heat
compound on the lamp’s surface with a clean, soft cloth or
similar material. Heat compound may cause the coating to
peel or the lamp to explode.
• Do not touch the lamp’s glass surface. Fingerprints may
cause the lamp to explode.
Lamp
Guide pin
Heat compound