EasyManua.ls Logo

OPTi CCA-TS2 - Page 141

OPTi CCA-TS2
299 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operator’s Manual – OPTI CCA-TS2 8-11
8 DIAGNOSTICS AND TROUBLESHOOTING
<ERROR! - Passwords Don’t Match>
The password entered was incorrect.
• Press
to continue.
• Enter the correct password.
<ERROR! - Permission Denied>
Permission denied since user does not have access
privileges for the selected function.
• Press
to continue. Information on
setting up user permissions can be found in
Section 3.2.3.1.3.
<Range Error>
This error may occur during Hb calibration.
The error is triggered, when the correction is greater
than 10%.
• Press
and replace the Hb calibrator.
<ERROR! - Not enough records in Level 2>
There are not enough records in the database to
generate a statistics report.
• Press
to continue.
<ERROR! - Group Name Already Exists>
The selected group name already exists in the
database.
• Press
to continue.
• Enter a unique group name.

Table of Contents