pH-B-2 Operator’s Manual – OPTI CCA-TS2
pH (Dry Sensor) ANALYTES
Standard Reference Cassette (SRC) Limit Values
LOW NORMAL HIGH Units
7.100 ± 0.02 7.400 ± 0.02 7.600 ± 0.02 pH units
Interferences
The following exogenous interferents were quantied in tonometered whole blood samples spiked with a
number of endogenous and exogenous chemicals and tested for interference following the CLSI guideline
EP7-A2, showing interferences to dyes which typically have short half-lives within the body before being
metabolized by the liver. This testing was done on the OPTI CCA-TS platform, with a standard OPTI
style pH sensor. No changes were made to the sensor or the measurement principle, so the interference of
the OPTI CCA-TS2 is equivalent to the OPTI CCA-TS
:
Chemical Interferent
Concentration
pH Level Interference
Acetaminophen 1.66 mM
7.170 NO
7.520 NO
Acetylsalicylic acid 3.33 mM
7.170 NO
7.520 NO
Ascorbic acid 0.23 mM
7.170 NO
7.520 NO
B-Hydroxybutyric acid 16.03 mM
7.170 NO
7.520 NO
Bilirubin 0.26 mM
7.170 NO
7.520 NO
Cardiogreen 0.0065 mM
7.170 NO
7.520 NO
Cystein 6.41 mM
7.170 NO
7.520 NO
Ethanol 86.8 mM
7.170 NO
7.520 NO
Evans blue 0.0104 mM
7.170 0.140
7.520 NO
Glycolic acid 10 mM
7.170 NO
7.520 NO
Halothane 0.759 mM
7.170 NO
7.520 NO
Ibuprofen 2.43 mM
7.170 NO
7.520 NO
Intralipid 1%
7.170 NO
7.520 NO