PCO2-B-2 Operator’s Manual – OPTI CCA-TS2
PCO
2
(Dry Sensor) ANALYTES
Interferences
The following exogenous interferents were quantied in tonometered whole blood samples
spiked with a number of endogenous and exogenous chemicals and tested for interference
following the CLSI guideline EP7-A2, showing interferences to dyes which typically have short
half-lives within the body before being metabolized by the liver. This testing was done on the
OPTI CCA-TS platform, with a standard OPTI style pCO
2
sensor. No changes were made to the
sensor or the measurement principle, so the interference of the OPTI CCA-TS2 is equivalent to
the OPTI CCA-TS.
Chemical Interferent
Concentration
PCO
2
Level Interference
Acetaminophen 1.66 mM
83 mmHg NO
17 mmHg NO
Acetylsalicylic acid 3.33 mM
83 mmHg NO
17 mmHg NO
Ascorbic acid 0.23 mM
83 mmHg NO
17 mmHg NO
B-Hydroxybutyric acid 16.03 mM
83 mmHg NO
17 mmHg NO
Bilirubin 0.26 mM
83 mmHg NO
17 mmHg NO
Cardiogreen 0.0065 mM
83 mmHg NO
17 mmHg NO
Cystein 6.41 mM
83 mmHg NO
17 mmHg NO
Ethanol 86.8 mM
83 mmHg NO
17 mmHg NO
Evans blue 0.0104 mM
83 mmHg -24.68 mmHg
17 mmHg NO
Glycolic acid 10 mM
83 mmHg NO
17 mmHg NO
Halothane 0.759 mM
83 mmHg NO
17 mmHg NO
Ibuprofen 2.43 mM
83 mmHg NO
17 mmHg NO
Intralipid 1%
83 mmHg NO
17 mmHg NO
Methylene Blue 0.125 mM
83 mmHg NO
17 mmHg NO
Sodium Chloride 20 mM
83 mmHg NO
17 mmHg NO