Na-4 Operator’s Manual – OPTI CCA-TS2
SODIUM ANALYTES
Direct linearity for the OPTI CCA-TS2 measurement has been established against reference materials or
methods. The linearity of the Sodium sensor on the OPTI CCA-TS2 has been established versus the same
sensor on the OPTI CCA-TS using whole blood samples tonometered with mid level mixtures and spiked
or diluted to establish the correlation.
Parameter and Specication Value
Number of Samples 63
Slope 1.00
Offset -0.18
R
2
0.999
Sy|x 0.781
80
100
120
140
160
180
200
220
80 100 120 140 160 180 200 220
Linear fit
(-0.18 + 1.00x)