Operator’s Manual – OPTI CCA-TS2 Cl-5
ANALYTES CHLORIDE
Correlation to Other Methods
3
OPTI CCA-TS vs other Electrolyte Instruments on whole blood in a typical setting
Excess blood aliquots from specimens collected for electrolyte analyses were analyzed by both traditional
and non-traditional operators of measurement equipment in hospital laboratories. The sample was
analyzed on the OPTI CCA-TS after obtaining the requisite results from existing instrumentation used for
these analyses and operated and controlled following their established procedures.
Correlation
Comparative Method* Slope Intercept Correlation
Coefcient
Sy.x Range n
Analyzer A (whole blood) 0.9965 0.95 0.9246 1.96 92 to 117 173
* For more information on specic analyzers used, please contact OPTI Medical Technical Support.