Measurement Principle
The glucose optode measurement is based on the enzymatic oxidation of glucose.
Glucose + O
2
gluconic acid + H
2
O
2
glucose oxidase
The sensor is constructed of an enzyme layer over an oxygen sensor. As a sample containing glucose
contacts the sensor
, the oxidation of the glucose consumes the oxygen locally present in the sensor.
This decrease in oxygen is detected in the same manner (luminescence quenching) as described for the
PO
2
optode. The amount of glucose is determined to be proportional to the rate at which the oxygen is
consumed.
Measurement Range
Range Resolution (Low/High) Units
30 to 400
(70 to 400 mg/dL for samples with
pO
2
levels from 401-700 mmHg)
0.1/0.1 mg/dL
1.7 to 22 0.01/0.01 mmol/L
Standard Reference Cassette (SRC) Limit Values
LOW NORMAL HIGH Units
40.0 ± 4.0 110.0 ± 4.0 300.0 ± 4.0 mg/dL
2.2 ±0.22 6.1 ±0.22 16.65 ±0.22 mmol/L
Interferences
The OPTI CCA Glucose sensor does exhibit interference from PO
2
levels that exceed 700 mmHg.
The Glucose sensor corrects for PO
2
values up to 700 mmHg. Glucose values are suppressed when PO
2
values are > 700 mmHg. This testing was done on the OPTI CCA platform, with a standard OPTI style
Glucose sensor. No changes were made to the sensor or the measurement principle, so the interference of
the OPTI CCA-TS2 is equivalent to the OPTI CCA.
Heparin salts are the only acceptable anticoagulants. Other anticoagulants such as citrate, EDTA, oxalate,
and uoride cause signicant interferences to the Glucose sensor.
Only clear, uncolored quality control materials, such as OPTI CHECK PLUS brand aqueous controls
should be used with the OPTI CCA-TS2 system. Colored materials, including prociency testing
materials, may interfere with the Glucose measurement, or fail to be properly aspirated.