EasyManua.ls Logo

OPTi CCA-TS2 - Page 256

OPTi CCA-TS2
299 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
BUN-6 Operator’s Manual – OPTI CCA-TS2
BUN/UREA ANALYTES
OPTI CCA-TS2 vs OPTI CCA-TS on whole blood in a typical setting
In Method Comparisons carried out following the experimental protocol recommended in the CLSI
guideline EP9-A2, Volume 15, Number 19 using whole blood samples on the OPTI CCA-TS2 versus the
OPTI CCA-TS.
Parameter and Specication Value
Number of Samples 45
Range 5.5 to 103.2
Slope 1.01
Offset -0.20
R
2
0.999
Sy|x 0.883
References
1. Tietz, Norbert W., Ed., ClinicalGuidetoLaboratoryTests, 2nd Ed., (Philadelphia: W.B.Saunders,
Co., 1990) p. 436
.
2. Tietz, Burtis C.(Ed.), TextbookofClinicalChemistryandMolecularDiagnostics, 4
th
Ed., (Elsevier
Saunders, 2006), p. 802.
3. OPTI Medical. Model equation for regression statistics is: [results of OPTI Analyzer] = slope(m)
[comparative method results] + intercept(b).
0
20
40
60
80
100
120
0 20 40 60 80 100 120
BUN - TS2
BUN - TS
Scatter Plot with Fit
Identity
Linear fit
(-0.20 + 1.01x)