Operator’s Manual – OPTI CCA-TS2 Lac-5
ANALYTES LACTATE
Parameter and Specication Value
Number of Samples 55
Slope 1.04
Offset -0.02
R
2
0.995
Sy|x 0.390
Correlation to Other Methods
2
OPTI CCA-TS vs other Lactate Instruments on whole blood in a typical setting
Excess sample aliquots from specimens collected for metabolyte analyses were analyzed by both
traditional and non-traditional operators of measurement equipment in hospital laboratories. The sample
was analyzed on the OPTI CCA-TS after obtaining the requisite results from existing instrumentation
used for these analyses and operated and controlled following their established procedures.
Correlation
Comparative Method* Slope Intercept Correlation
Coefcient
Sy.x Range n
Analyzer A (whole blood) 0.96 -0.01 0.944 0.64 0.78 to 13.96 175
Analyzer B (whole blood) 1.18 -0.54 0.953 0.55 0.40 to 11.3 49
* For more information on specic analyzers used, please contact OPTI Medical Technical Support.
0
2
4
6
8
10
12
14
16
18
20
0 2 4 6 8 10 12 14 16 18 20
Linear fit
(-0.02 + 1.04x)