Operator’s Manual – OPTI CCA-TS2 A-1
APPENDIX A - TECHNICAL SPECIFICATIONS
APPENDIX A - TECHNICAL SPECIFICATIONS
Measurement Range
Parameter Range Display Units
Resolution
(Lo/Hi)
pH 6.6 to 7.8 0.01/0.001 pH units
PCO
2
10 to 200 1/0.1 mmHg
PO
2
10 to 700 1/0.1 mmHg
Na
+
100 to 180 1/0.1 mmol/L
K
+
0.8 to 10 0.1/0.01 mmol/L
Ca
++
0.2 to 3.0 0.01 mmol/L
Cl- 50 to 160 1/0.1 mmol/L
Glu 30 to 400 0.1 mg/dL
(70 to 400 for samples with PO
2
levels between 401-700 mmHg)
Glu 1.7 to 22 0.01 mmol/L
BUN 2.8 to 112 0.1 mg/dL
Urea 1 to 40 0.01 mmol/L
Lac 0.3 to 17.5 0.01 mmol/L
tHb 5 to 25 0.1 g/dL
SO
2
60 to 100 1/0.1 %
Barometric Pressure
300 to 800 mmHg
Operating Altitude
Up to 3048m (10,000ft)
Pollution Degree
Degree 2, normal indoor laboratory environment. Air contains only non-conductive
pollutants with occasional condensation.