EasyManua.ls Logo

OPTi CCA-TS2 - Page 50

OPTi CCA-TS2
299 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3-14 Operator’s Manual – OPTI CCA-TS2
3 CUSTOMIZATION
For information on how to change units of
measure, see section 3.2.4.2.
NOTE: Eachfacilityshouldestablishitsown
referenceranges.Thepresetanalyzerranges
areforreferenceonlyandarederivedfrom
“Tietz,BurtisC,etal(Eds.),Textbook
ofClinicalChemistryandMolecular
Diagnostics,4thEd.,(ElsevierSaunders,
2006)pps.2252-2302”.
6. Select the <Critical> tab (Fig. 3-28).
By default, critical ranges are disabled.
Press <Enable> to enter critical limit values.
7. Select the parameter you want to change and
press to enter the new limit value.
8. Press to access pages 2 and 3 with the
remaining parameters (Fig. 3-29 and Fig. 3-30).
9. Press to accept the new limit values.
The instrument is preset to the measurement ranges
of the OPTI CCA-TS2:
pH: 6.600 - 7.800
PCO
2
: 10.0 - 200.0 mmHg
PO
2
: 10.0 - 700.0 mmHg
Na
+
: 100.0 - 180.0 mmol/L
K
+
: 0.80 - 9.99 mmol/L
Cl
-
: 50.0 - 160.0 mmol/L
Ca
++
: 0.20 - 3.00 mmol/L
Glu: 30.0 - 400.0 mg/dL
Glu: 1.7 - 22.0 mmol/L
BUN: 2.8 - 112.0 mg/dL
Urea: 1.0 - 40.0 mmol/L
Lac: 0.30 - 17.50 mmol/L
tHb: 5.0 - 25.0 g/dL
SO
2
: 60.0 - 100.0 %
In all data input screens, if values outside the
acceptable input range are entered, the system
automatically ags the error and displays the
valid range.
The limits entered here will reside in the
instrument memory even after system power is
turned off.
10. Press to return to the <Setup> screen or
to return to the main menu.
Fig. 3-28 Enter Critical Limits
Fig. 3-29 Critical Limits 2
Fig. 3-30 Critical Limits 3

Table of Contents