5-8 Operator’s Manual – OPTI CCA-TS2
5 SAMPLE HANDLING AND PATIENT TESTING
7. The system will now check the integrity of the
cassette and then calibrate (Fig. 5-9). For more
information about calibration, please refer to
Chapter 4 “Calibration and Quality Control”.
NOTE: TheOPTICCA-TS2willholdcalibration
for10minutesforallcassettetypesexcept
B-Lac.B-Laccassetteswillholdcalibration
for2minutes.TheOPTICCA-TS2willbeep,
warningyouwhenonly1minuteremains
onthecalibration.Afterthistimeelapses,
amessagewillbedisplayedtodiscardthe
cassette.
NOTE: IftHb/SO
2
hasbeendisabled(seeSection
3.2.2.2), youmayattachthesampleatany
timeduringcalibrationandpress .
Thesamplewillthenbeautomatically
aspiratedaftercalibrationandthe
measurementwillbegin.
8. You can enter patient information while
calibration is in progress by pressing <Patient
Info> (Fig. 5-10).
•
Press the <Last Patient Info> button
(Fig. 5-10) to use the last patient info as the
default for the current patient information,
if this option is enabled in <Setup> (see
Section 3.2.2.2).
•
This option will populate all patient info
elds with the last patient data.
If user ID security is enabled, the user ID
eld will contain the user ID of the user
currently logged in. All patient information
used as the default can be edited.
• Verify that patient ID and all other input
parameters are correct for every patient
sample measurement.
•
Press the <Patient Info> button to enter
new patient information or to not use the last
patient info as the default.
Fig. 5-9 Cassette Calibration
Fig. 5-10 Patient Info