5-8 Operator’s Manual – OPTI CCA-TS2
5 SAMPLE HANDLING AND PATIENT TESTING
7. The system will now check the integrity of the
cassette and then calibrate (Fig. 5-9). For more
information about calibration, please refer to
Chapter 4 “Calibration and Quality Control”.
NOTE: TheOPTICCA-TS2willholdcalibration
for10minutesforallcassettetypesexcept
B-Lac.B-Laccassetteswillholdcalibration
for2minutes.TheOPTICCA-TS2willbeep,
warningyouwhenonly1minuteremains
onthecalibration.Afterthistimeelapses,
amessagewillbedisplayedtodiscardthe
cassette.
NOTE: IftHb/SO
2
hasbeendisabled(seeSection
3.2.2.2), youmayattachthesampleatany
timeduringcalibrationandpress .
Thesamplewillthenbeautomatically
aspiratedaftercalibrationandthe
measurementwillbegin.
8. You can enter patient information while
calibration is in progress by pressing <Patient
Info> (Fig. 5-10).
•
Press the <Last Patient Info> button
(Fig. 5-10) to use the last patient info as the
default for the current patient information,
if this option is enabled in <Setup> (see
Section 3.2.2.2).
•
This option will populate all patient info
elds with the last patient data.
If user ID security is enabled, the user ID
eld will contain the user ID of the user
currently logged in. All patient information
used as the default can be edited.
• Verify that patient ID and all other input
parameters are correct for every patient
sample measurement.
•
Press the <Patient Info> button to enter
new patient information or to not use the last
patient info as the default.
Fig. 5-9 Cassette Calibration
Fig. 5-10 Patient Info