EasyManua.ls Logo

OPTi CCA-TS2 - Page 97

OPTi CCA-TS2
299 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operator’s Manual – OPTI CCA-TS2 5-7
5 SAMPLE HANDLING AND PATIENT TESTING
5. Press down on the red latch to open the Sample
Measurement Chamber (SMC) (Fig. 5-6).
6. Insert the cassette as follows:
• Open the OPTI Sensor Cassette pouch and
remove the cassette (Fig. 5-7). After opening
the pouch, proceed with the following steps
immediately.
NOTE: Thecassetteshouldberunimmediately
afteropeningthepouch,butnolaterthan2
minutes(1minuteforB-LacCassettes)after
opening.
NOTE: Forsampleintroductionwithacapillary
tubeoraComfortSampler,removethe
syringeadapterbeforeplacingthecassette
intothechamber.
• Gently wipe both sides of the cassette with a
clean dry cloth to remove excess moisture.
• Insert the cassette in the chamber.
Press down to ensure the cassette is properly
seated (Fig. 5-7).
• Close the SMC cover by pressing it down
rmly (Fig. 5-8).
• The green status light starts to blink
indicating that the SMC cover should not be
opened during this time.
NOTE: IftheSMCcoverisopenedwhilethe
greenstatuslightisblinking,thecassette
calibrationwillbecancelledandthe
cassettemustbediscarded.
Fig. 5-7 Insert Cassette
Fig. 5-6 Open SMC Cover
Fig. 5-8 Close Cover

Table of Contents