User Manual for Otoport DP+TE model
129
Conguration
15.2.6 ABR
The Otoport OAE+ABR has two di󰀨erent ABR test modes, ABRscreen and
ABRcustom. ABRscreen is optimised for the screening of neonates and
is less congurable than ABRcustom. Choose the mode you wish to edit
from the menu.
Stim level
‘Stim Level’ sets the stimulus level used in the ABR test in dBHL.
If Stim Level is set to OFF then the selected mode (ABRscreen or
ABRcustom) is not available for testing and it will not appear in the main
menu carousel.
The ‘ABRscreen mode is intended for newborn hearing screening.
The stim level can be set from 30-45dBHL in 5dB steps.
The default setting is 40dBHL, but you should be set this level according to
your screening program.
The stimulus level should be set to the lowest value compatible with
the aims and targets of the hearing screening program. For the reliable
detection of mild hearing impairment many professional bodies recommend
stimulus levels 35dBnHL (e.g. American Speech Language Hearing