DPOAE test
8.4 DPOAE test results
The following table lists all possible test results with the associated results
graphic and gives an explanation of the circumstances under which each
result would be shown.
Test Result Description
DPOAE Pass The data collected has met the criteria set. The
optimum test setting will depend on your application,
e.g. screening or clinical measurement
(see
15.2 Test settings)
Note: The following results will only occur if a DPOAE Pass is not obtained,
providing test information feedback to the user
No Valid OAE The data collected has not met the set pass criteria
and the test conditions were acceptable
Noisy The noisy data collected is three times greater than
the low noise data collected
Poor Probe Fit Probe t movement is detected that results in less
than 85% of stimuli reaching calibration levels
Too Few Bands Insu󰀩cient bands meet their pass criteria
Stopped Too The test has been ended manually before the
Soon required minimum amount of data has been collected
Select Save to skip to the Patient Details screen ready to Save.
Select Review to view the data of the completed test, prior to saving.
Select Cancel to discard the test result. A prompt requiring conrmation
will be shown before the result is discarded.