CHAPTER FIVE
Test preparation
52
Test preparation
5 Test preparation
Good preparation for testing will improve test results and make testing
more e󰀩cient and less stressful for the baby, parents and tester.
5.1 General checks before testing
Ensure the Otoport is charged (see chapter 21 Otoport power for
information).
Ensure the Otoport weekly checks are being regularly conducted (see
chapter 17 Quality checks for information).
Do not run a test if there is any discharge from the ear to be tested.
Choose a quiet room, without background noises.
Ensure the patient is comfortable and settled.
Ensure you can clearly see the ear to be tested.
5.2 Environment checks for ABR
Try to minimise any interference from electrical equipment. Turn o󰀨
electrical equipment and orescent lighting if possible. If equipment cannot
be turned o󰀨, try to move away from it.
Ensure that all necessary disposables (sensors, tips etc) are at hand and
prepared for use.
Plan where the baby will be positioned, where you will place the Otoport so
it will be visible and secure and how you will route the electrode and probe
cables.