User Manual for Otoport DP+TE model
177
Care of the Otoport
Before tting each tip, ensure the sound tubes are carefully examined for
any sign of debris that may have entered them. Replace any part of the
probe ear piece as necessary. (See chapter 19 Probe care for details)
Ensure your hands are cleaned thoroughly between each patient tested.
Clean the Otoport each day before a testing session, or according to local
requirements. Ensure the Otoport is cleaned if it becomes contaminated.
Clean surfaces of the Otoport with an alcohol based sterile wipe or cloth
and antiseptic uid. Dry the Otoport with tissue immediately afterwards. Do
not allow liquid to enter the instrument and do not immerse in uid. Do not
allow liquid to come into contact with the connection sockets. Do no poke
any materials into either the probe or charger/pc cable sockets.
If additional hygienic protection is required, use the Otoport in an infection
control sleeve. This can also be cleaned with a sterile wipe or cloth with
antiseptic uid. The sleeves are disposable, so should be replaced weekly
or after approximately every 50 tests.