Span
The system can output measurement values with a span (factor) multiplied.
Final measurement value = Span x measurement value + offset
Setting range Function Initial value
0.1000 to
+9.9999
Set a span in a range from
0.1000 to +9.9999.
The display limit of the measurement value is ±95.000. Make sure that the setting
value does not exceed the display limit.
Ɣ6HWWLQJ3URFHGXUH
1
Display the "Data Processing Setting"
screen while the system is in setting mode.
2
3UHVVWKH>(17(5@NH\7KH$YHUDJH
Time" screen will be displayed.
3
3UHVVWKH>'2:1@NH\WZLFH7KH6SDQ
screen will be displayed.
4
PreVVWKH>(17(5@NH\7KHSUHVHQW
span will be displayed. The leftmost digit
will be selected.
5
3UHVVWKH>83@>'2:1@NH\VWRJHWKHUWR
change the selected digit. After all the
FKDQJHGGLJLWVKDYHEOLQNHGWKHVLJQFDQ
be changed. For details, refer to page 3-8
0DNLQJ6HWWLQJ9DOXH&KDQJHV
Output
Final meas. value
Meas. value
Spanned meas. value
Offset value
Measurem
ent range
0
Center point of measurement