3-3-7㻌Alarm Settings
Analog Output at Alarm㻌
This function is used to set the analog output status at alarm operation.
Analog output in the alarm state (where measurement is disabled owing to lack of
light) can be kept on hold or set to a fixed value selectively.
㻌
㻌
Setting Function Panel display
Hold
Holds the analog output
immediately before the alarm.
㻌
㻌
Fixed Value
Fixed-value output
㻌
㻌
* indicates the default value.
㻌
㻌
㻌
A fixed current output of 21.6 [mA] or voltage output of +11.000 [V] applies.
㻌
㻌
The fixed value is applied to the side set with "Analog Output Selection."
㻌
㻌
㻌
Check the input range of your input device such as AD board before setting.
㻌
㻌
The above settings will apply when there is data unstability.
㻌
Ɣ6HWWLQJ3URFHGXUH
㻌
1
Display the "Alarm Setting" screen while
in setting mode.
㻌
2
3UHVVWKH>(17(5@NH\7KH$QDORJ
output at alarm" will be displayed.
㻌
3
3UHVVWKH>(17(5@NH\7KH+ROG)L[
will be displayed.
㻌
4
3UHVVWKH>83@>'2:1@NH\WRVHWWKe
system to Hold/Fix. (Example: Changing
the selection to the fixed value.)
㻌
5
3UHVVWKH>(17(5@NH\WRGHWHUPLQHWKHVHW
value.