4
GettinG Started
Pinnacle Operators Manual
Pickering Laboratories Inc.
Radio Frequency Pinnacle PCX complies with IEC 61000-4-3
This device complies with Part 15 CHECK Class B of the FCC Rules. Operation is subject to the following
two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any
interference received, including interference that may cause undesired operation.
WETTED MATERIALS
• PEEK, PTFE, Air Barrier Tubing
• 99.9% Ceramic
• Perlast
®
• Borosilicate glass
• EPR
SOFTWARE
PCX Control Software runs on Microsoft Windows XP, Vista, or Windows 7.
10Mb hard disk space
PCX History Log file: Maximum file Size on the computer: 1.2 Mb
Site Requirements
INSTRUMENT
Bench Space
33 H x 17 W x 21 D inches (84 x 56 x 59 cm), both doors fully opened, with bottles and electrical connections
in place.
Minimum 3 inches clearance at back of instrument for venting.
Note: Space quoted above is for Pinnacle PCX.
The total space requirement depends on the brand and model of HPLC.
Electrical Outlet
One grounded outlet must be provided for the Pinnacle PCX.