3.4
Section 3
InstallatIon
Pinnacle Operators Manual
Pickering Laboratories Inc.
FOR CARBAMATE ANALYSIS
Pump
Minimum binary gradient elution
Injector
For water samples, at least 200μl injection
For all other applications, review the method notes for chemistry requirements.
HPLC RELAY REQUIREMENTS
For HPLC systems other than Agilent 1100, or 1200, the Software and system must be capable of sending a relay
signal to an external piece of equipment to achieve synchronization.
Chemstation version 9.0 or higher is needed for Agilent 1100 or Agilent 1200. Pinnacle PCX software will
communicate with Chemstation directly – no relay connection is needed.
Any machine that drives this relay input shall provide a relay contact pair that is electrically isolated from all
other electrical devices. The relay signal must have:
Relay detection voltage 24 +/- 2 V
Relay detection current Approximately 1 mA
COMPUTER REQUIREMENTS
For the Installation, we strongly recommend installing the USB cable.
Use of the network connection is optional at the user’s discretion.
The computer must have:
Microsoft Windows XP, Vista, and Windows 7.
Minimum of one extra USB port for installation
Available Memory: Minimum 10Mb
Instrument Unpacking and Preparation
UNPACKING
Unpack all cartons and review the contents using the Packing List to ensure that all of the items are present. If
any items are missing, immediately contact Pickering Laboratories, Inc.