EasyManua.ls Logo

Pickering Laboratories Pinnacle PCX - Pump and Autosampler Connections

Default Icon
179 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3.8
Section 3
InstallatIon
Pinnacle Operators Manual
Pickering Laboratories Inc.
Press MODE button. The message will read:
RUN MAIN F/W
DHCP IP: 192.168.111.111
Follow the chart below to make changes in IP address.
Restart the instrument after completing the steps above.
Pump and Autosampler Connections
Note: Do not fit the analytical column and guard yet.
Replace any mobile phase that is more than 2 weeks old, especially DI water. Rinse the reservoir bottles
thoroughly with soap and water. Wipe down the dip tubes on the caps with methanol and a clean, lint-free
cellulose tissue. Avoid touching the tubing or the interior of the reservoir with your skin and do not leave caps
and lines dangling without a reservoir because this can cause contamination.
Fill reservoirs with 80/20 Water/Methanol
Open the prime-purge valve on the HPLC. Purge each line in the system with a water/methanol mixture to flush
the system. Set the flow rate to the maximum and purge at least 25 ml through each line.
Main LCD Display
DHCP IP: 152.168.111.111
NETWORK
IP FROM DHCP
DHCP IP: 192.168.111.111
MAC: 00:50:C2:44:30:45
PRESSURE UNITS
PSI
RUN MAIN F/W
DHCP IP: 192.168.111.111
NETWORK
STATIC IP
STATIC IP:192.168.111.111
MAC:00:50:C2:44:30:45
EDIT IP: 192.168.111.111
Select changes number
Mode moves next to position
PRESSURE UNITS
BAR
Select
Select
Select
Select
Mode
Mode
Mode
Mode
Mode

Table of Contents