3.16
Section 3
InstallatIon
Pinnacle Operators Manual
Pickering Laboratories Inc.
RECOMMENDED ACCOMPANYING HPLC SLOWDOWN METHOD
Set the HPLC to 100% Storage Eluant (see application section for proper eluant for your column), and set the
HPLC pump at the normal analytical flow rate. Choose an eluant that elutes contaminants from the column; for
example, methanol for a reversed-phase column and regenerant for an ion-exchange column.
Time (min) % Storage Eluant Flow (ml/min)
0 100 Analytical flow rate*
30 100 Analytical flow rate*
30.1 100 0.0 or 0.02mL/min
*Follow instructions that come with the column.