Check the operation of the safety switches
D
elivery arm
Check the friction of the delivery arm joints
Adjust, if needed.
Check the balance of the delivery arm
Adjust, if needed.
Check the instrument hoses and quick connectors
Check the function of the control panel
Check the operation of the syringe
Adjust air/water flow from the syringe multiplexer block, if needed.
Check the operation of the turbine
Pressure: __________bar.
Check the operation of the micromotor
Check the operation of the polymerisation light
Check the operation of the ultrasonic scaler
If with LED, check that service mode 39 is set to maximum.
P
atient chair
Visually check the condition of the chair and upholsteries
Check the calibration of motors in service modes 79 & 80
Recalibrate, if necessary.
Check the programming function and general operation
Check the operation of the emergency switches
Backrest, seat and lifting arm.
Check the operation of the headrest locking mechanism
S
olanna operating light
Check the mechanical installation
22 Post-installation checklist