15 Attaching tray table assembly to OP delivery arm
balanced arm instrument console
The mounting arm is attached to the instrument console by pushing it to its
position.
The friction of the tray arm movements can be adjusted. Loosen the joint
attachment screw using a 3 mm Allen key (1). Rotate the plate manually and
move the tray arm (2). Hold the plate in position with pliers (3) and tighten
the attachment screw (4).
The angle of the tray table can be adjusted as follows. Loosen the securing
screws (5) and adjust the angle (7) with four adjustment screws (6). Tighten
the securing screws.
15 Attaching tray table assembly to OP delivery arm balanced arm instrument console
80 Planmeca Compact i3 Installation manual