Attach the grounding lead (2) to the monitor grounding point through the
handle opening with a screw (1) and serrated washer (3) using a 3 mm Allen
key. Connect the power cable to the monitor connector (4) Connect the
HDMI cable to the monitor connector (5).
10 Installing monitor