4.e. Connect the assistant syringe water tube to the manifold.
Remove the plug and replace it with the manifold nipple. Attach
the assistant syringe water tube and manifold bush to the nipple.
Secure the tube into place with the bush.
4.f.
Connect the assistant syringe water tube that comes from the
manifold (7) to the assistant syringe valve.
4.g. Connect the water tube that goes to the manifold, instrument
console and clean-water bottle (DCI) (7) to the water filter/
regulator.
4.h. Connect the air tube that goes to the manifold, instrument console
and clean-water bottle (DCI) (7) to the air filter/regulator.