Note: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to
part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference
when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate
radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful
interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful
interference in which case the user will be required to correct the interference at his own expense.
This ISM device complies with Canadian ICES-001. (Cet appareil ISM est conforme à la norme NMB-001 du Canada.)
Warning: This is a Class A product. In a domestic environment this product may cause radio interference in which
case the user may be required to take adequate measures.
Guidance and Manufacturers Declaration - Electromagnetic Emissions
The Emerald is intended for use in the electromagnetic environment specied below. The customer or the user of the
Emerald should assure that it is used in such an environment.
Emissions Test Compliance Electromagnetic Environment - guidance
RF emissions CISPR 11 Group 1 The Emerald uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
RF emissions CISPR 11 Class A The Emerald is suitable for use in all establishments other
than domestic and those directly connected to the public
low voltage power supply network that supplies buildings
used for domestic purposes.
Harmonic emissions IEC 61000-3-2 Not applicable
Voltage uctuations / icker
emissions IEC 61000-3-3
Not applicable
Guidance and Manufacturers Declaration - Electromagnetic Immunity
The Emerald is intended for use in the electromagnetic environment specied below. The customer or the user of the
Emerald should ensure that it is used in such an environment.
Immunity Test IEC 60601 Test Level Compliance
Level
Electromagnetic Environment -
guidance
Electrostatic discharge
(ESD) IEC 61000-4-2
+/- 6 kV contact
+/- 8 kV air
+/- 6 kV contact
+/- 8 kV air
Floors should be wood, concrete
or ceramic tile. If oors are covered
with synthetic material, the relative
humidity should be at least 30%
Electrical fast transient/
burst IEC 61000-4-4
+/- 2 kV for power supply lines
+/- 1 kV for input/output
Not applicable Mains power quality should be that
of a typical commercial or hospital
environment
Surge
IEC 61000-4-5
+/- 1 kV dierential mode
+/- 2 kV common mode
Not applicable Mains power quality should be that
of a typical commercial or hospital
environment
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
<5% Ut
(>95% dip in Ut)for 0,5 cycle
40% Ut
(60% dip in Ut) for 5 cycles
70% Ut
(30% dip in Ut) for 25 cycles
<5% Ut
(>95% dip in Ut)for 5 sec.
Not applicable Mains power quality should be that
of a typical commercial or hospital
environment.
15698200.B Planmeca Emerald Specications 49