Recommended separation distances between portable and mobile RF communications equipment and the
Emerald
For transmitters rated at a maximum output power not listed above, the recommended separation distance d
in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is aected by absorption and
reection from structures, objects and people.
Emerald Optical Specications
Caution Using controls, making adjustments, or performing procedures in a manner other than is specied in this
documentation may result in hazardous radiation exposure.
Laser Class
a
Class 2
Output 6.2 mW
Wavelength Range 400 - 700 nm
Beam divergence 10 degrees
a. Laser product classied to standard IEC/EN 60825-1:2007-03 Ed. 2.0
The scanner’s laser projection system utilizes a divergent beam powered by a non-accessible laser source with
a maximum power output of 200 mW. The scanner incorporates design features that prevent exposure to any
hazardous levels of laser radiation in normal operation modes and in any reasonable fault conditions.
52 Planmeca Emerald Specications Planmeca Emerald User Manual