Loading...
Immunity Test IEC 60601 Test
Level
Compliance
Level
Electromagnetic
Environment - guidance
Surge
IEC 61000-4-5
+/- 1 kV
dierential mode
+/- 2 kV common
mode
Not
applicable
Mains power quality
should be that of a typical
commercial or hospital
environment
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
<5% Ut
(>95% dip in Ut)
for 0,5 cycle
40% Ut
(60% dip in Ut) for
5 cycles
70% Ut
(30% dip in Ut) for
25 cycles
<5% Ut
(>95% dip in Ut)
for 5 sec.
Not
applicable
Mains power quality
should be that of a typical
commercial or hospital
environment.
Power
frequency
(50/60 Hz)
Magnetic eld
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic
elds should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Note: Ut is the a.c. mains voltage prior to application of the test level.