Recommended separation distances between portable and mobile RF
communications equipment and the PlanScan
Rated
maximum
output power
of transmitter
W
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in metres (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the maximum
output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is aected by absorption and reection from structures, objects and
people.
PlanScan Optical Specications
Caution Using controls, making adjustments, or performing procedures in a manner other than
is specied in this documentation may result in hazardous radiation exposure.
Laser Class
a
Class 2
Output 4.95 mW
Wavelength 450 nm
Beam divergence 10 degrees
a. Laser product classied to standard IEC/EN 60825-1:2007-03 Ed. 2.0
The scanner’s laser projection system utilizes a divergent beam powered by a non-accessible
laser source with a maximum power output of 200 mW. The scanner incorporates design
features that prevent exposure to any hazardous levels of laser radiation in normal operation
modes and in any reasonable fault conditions.
52 Scanner Specications Planmeca PlanScan User Manual