Code Explanation Comments
H141 Height movement Movement stopped because the
column is moving in wrong direction.
The column is moving
in wrong direction, check
sensors and cables.
H142 Height movement is not possible
because the stop plate at the bottom
of the column was activated.
Clear any obstruction before
moving the column again.
H146 Motor safety switch in upper
direction z drive
H147 Motor safety switch in lower direction
z drive
H148 Height movement is not possible.
The position of the patient support
base is too high.
Use the height adjusting
slider to move the patient
support base down.
H149 Height movement was stopped
because the C-arm cannot be
moved higher.
Clear any obstruction before
continuing.
H150 Height movement was stopped
because the patient support base
cannot be moved lower.
Clear any obstruction before
continuing.
H151 Line voltage The line voltage was too low during
exposure.
Exposure was interrupted.
Contact your service
technician for help.
H152 The line voltage is too low. Exposure is not possible.
Contact your service
technician for help.
H153 Height movement Z column is too high.
H154 Z column is too low.
H155 Imaging equipment is too high.
H156 Imaging equipment is too low.
H157 Imaging equipment movement
timeout.
H158 Imaging equipment position sensor
not working properly.
H159 Z column position sensor not
working properly.
H160 Imaging equipment is moving in
wrong direction, check sensors and
cables.
14 Help messages
94 Planmeca Viso User's manual