Code Explanation Comments
H161 Temperature The temperature of the tube head is
too high.
Wait for a few minutes for the
tube head to cool down.
H162 The temperature of the lift motor is
too high.
Wait for a few minutes for the
lift motor to cool down.
H165 The temperature of the tube head is
too high for the selected exposure
values.
Wait for a few minutes for the
tube head to cool down.
H166 The maximum tube head energy
level was exceeded.
Wait for a few minutes for the
tube head to cool down or
use lower exposure values.
H170 User related
messages
Wrong licence code. Check the licence code.
H171 Timeout in production test.
H181 The imaging process was cancelled
in Planmeca Romexis.
H182 Timeout in image data transmission. Exposure was interrupted.
Contact your service
technician for help.
H186 No IP address defined for 3D
sensor.
H187 Problem during image data
transmission.
Exposure was interrupted.
Contact your service
technician for help.
H188 Reconstruction PC error Exposed image data exists
in reconstruction PC with the
dataset ID given in the error
message.
The image can be
brought into the Planmeca
Romexis with the Redo 3D
reconstructions command.
H189 The screen was touched during
exposure.
Exposure was interrupted.
H190 Protouch-CPU communication
failure.
H191 3D sensor communication failure.
H192 Workstation communication failure.
H193 Invalid scan settings.
H194 CPU connection not established.
H195 Request timed out while waiting for
CPU to respond.
H196 Version mismatch in communication
interfaces.
H197 Workstation communication
disabled.
H199 Video streaming failed.
14 Help messages