Page 88 of 110 150821/A4 Interface
3.1.5.34 SOURce:FREQuency[:CW]
Instruction: SOURce:FREQuency[:CW]
Parameter: Test Frequency
Unit: [HzkHz], default is Hz
Return Value: {Test Frequency} in <NR3> format
Function: Set or query the test frequency.
Range Values: 50Hz, 60Hz, 100Hz, 120Hz, 1kHz, 10kHz, 20kHz, 40kHz, 50kHz & 100kHz
3.1.5.35 SOURce:VOLTage[:LEVel][:IMMediate][AMPLitude]
Instruction: SOURce:VOLTage[:LEVel][:IMMediate][AMPLitude]
Parameter: Test Voltage
Return Value: {Test Voltage} in <NR3> format
Function: Set or query the test voltage.
Range Values: 0.02 ~ 1Vrms, Resolution is 10mV
3.1.5.36 SOURce:VOLTage[:LEVel][:IMMediate]:OFFSet:STATe
Instruction: SOURce:VOLTage[:LEVel][:IMMediate]:OFFSet:STATe
Parameter: {ON (1)OFF (0)}
Return Value: {10}
Function: Enable/Disable the DC bias voltage function.
Description: ON (1) Enable bias voltage
OFF (0) Disable bias voltage
3.1.5.37 SYSTem:ALARm
Instruction: SYSTem:ALARm
Parameter: {PULSeCONTinuous}
Return Value: {PULSeCONTinuous}
Function: Set the mode the alarm will sound in.
3.1.5.38 SYSTem:BEEPer:STATe
Instruction: SYSTem:BEEPer:STATe
Parameter: {OFF (0)ON, LOW (1)ON, HIGH (2)}
Return Value: {012}
Function: Set the loudness of the beeper.
3.1.5.39 SYSTem:CONSt
Instruction: SYSTem:CONSt
Parameter: Constant Source Resistance
Return Value: {100/251002510c}
Function: Set the constant source resistance.