Interface 150821/A4 Page 89 of 110
3.1.5.40 SYSTem:ERRor?
Instruction: SYSTem:ERRor?
Parameter: None
Return Value: Error message
Function: Query if there are any system errors.
3.1.5.41 SYSTem:HANDler
Instruction: SYSTem:HANDler
Parameter: {CLEArHOLD}
Return Value: {CLEArHOLD}
Function: Set the Handler to clear result or hold result for each test.
3.1.5.42 SYSTem:INTEgration
Instruction: SYSTem:INTEgration
Parameter: {0 - 8}
Return Value: {012345678}
Function: Set the integer cycle (# of sine waves integrated together per measurement).
3.1.5.43 SYSTem:KLOCk
Instruction: SYSTem:KLOCk
Parameter: {ON (1)OFF (0)}
Return Value: {10}
Function: Set or query if the Key Lock function is ON or OFF.
Description: ON (1) Turn on Key Lock
OFF (0) Turn off Key Lock
3.1.5.44 SYSTem:PRESet
Instruction: SYSTem:PRESet
Parameter: None
Return Value: None
Function: Set the instrument to initial default values.
3.1.5.45 SYSTem:VERSion
Instruction: SYSTem:VERSion
Parameter: None
Return Value: None
Function: Set the instrument software version and date.