Calibration Calibrating the Ultrasound Sensors
CubiScan 110-T 31
4. The following screen is displayed. Place the calibration weights on the
CubiScan 110-T platform.
5. Tap
[Next] to continue, and the following screen is displayed.
6. You have now finished calibrating the scale. Tap
HOME to return to the
home screen or if you would like to try calibrating the scale again, tap
[Next].
Calibrating the Ultrasound Sensors
To perform the calibration, you will need the following:
12'' x 5'' x 3.6'' calibration cube, supplied with the CubiScan
To calibrate the sensors using the touchscreen, proceed as follows.
Figure 24
Second Scale Calibration Screen
Figure 25
Scale Calibration Complete