5.9 Connecting Items and Tools
Replacements for any of these items are available from Repligen.
Contact Repligen at US +1 908-707-1009 or analytics-support@repligen.com
Figure 10. Connecting Items and Tools
5.10 Cary 60 Spectrophotometer and Accessories
Agilent Technologies’ Cary 60 spectrophotometer uses a Xenon flash lamp light source, which provides a wavelength range of
190 nm–1100 nm. Using the FlowVPX with the Cary 60 light source requires a Fiber Optic Coupler, provided by Repligen, to ensure
adequate light transmission through the Delivery Fiber to the FlowVPX instrument.
Note: Please reference documentation included with the Cary 60 spectrophotometer for component-specific requirements,
instructions, and safety considerations.
5.10.1 The FlowVPX Delivery Fiber
If using the Cary 60 spectrophotometer, the FlowVPX System requires a fiber optic cable to transmit the light from the Cary 60 to
the Flow Fibrette. This is called the FlowVPX Delivery Fiber. The Delivery Fiber included with the FlowVPX System is 3 m long; a
longer Delivery Fiber is available upon request.
CAUTION: Avoid bending the Delivery Fiber past a two-inch radius to ensure integrity of internal optical fiber.