EasyManua.ls Logo

REPLIGEN CTech FlowVPX System - Figure 50. Transmission Check Instructions (Cary 60)

Default Icon
55 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
40
DOC0211
eRev. 1.08/15/2023
User Guide
CTechFlowVPX® System
Figure 50. Transmission Check Instructions (Cary 60)
7. Load the Detector into the Detector Validation Adapter (DVA) unit.
Load the FlowVPX Detector into the DVA unit.
Ensure the Detector connector is facing out.
8. Connect the Delivery Fiber to the Transmission Tool.
Ensure the key on the fiber connector lines up with the keyway from the Transmission Tool’s FC adapter.
9. Connect Transmission tool to DVA.
Place the Transmission Tool onto the DVA with the assistance of the magnets.
Enter the wavelength(s) in the fields provided, which will determine at which wavelength(s) the system will take a
reading.
10. Please wait while ViPER runs a transmission test for each wavelength entered.
11. Click Finished once the transmission test(s) are complete.
ViPER will display the % Transmission as well as the Pass/Fail result of the test.
User has the option to retry the check if the %T value is not within acceptable range.

Table of Contents