DOC0211
eRev. 1.08/15/2023
User Guide
CTechFlowVPX® System
13. How to Run a System Suitability Test (Cary 60 or Beams)
Note: The ViPER software must first be installed and connected to successfully run a System Suitability test.
1. Run the ViPER software on the computer by double-clicking the appropriate program icon for the connected light source,
titled ViPER (Beams) or ViPER (Cary 60), located on the desktop.
2. Enter your username and password credentials in the indicated fields.
3. Click ValidateVPT to launch application.
Figure 54. My Applications: Validate VPT App
4. Flow Cell must be unloaded to begin.
Please ensure that no sample is flowing through the cell before unloading the Flow Cell. This may cause damage to
the System and to the surrounding environment.
Note: The FlowVPX instrument must be uninstalled from the Standard Mount.
5. Click the System Suitability button on the Instrument Control to begin the process.
Figure 55. System Suitability Button in Instrument Control Menu
13.1 Running a System Suitability Test Using a Cary 60 Light Source
If using the Cary 60 light source, complete the System Suitability Test by completing the following section. If using the Beams light
source, skip to section 13.2.