21.3.9 BUZ: Buzzer
Enable/disable the buzzer.
General
No. of parameters 1
Save changes With TDD1
Increment Trade Counter No
Parameter Details
Parameter Description Range Default
1 Enable 0 = Off
1 = On
1
Example
Command Answer Description
S01; Select unit 1
BUZ?; 1 CRLF Query setting, buzzer currently enabled
BUZ0; 0 CRLF Buzzer disabled
TDD1; 0 CRLF Save new settings
21.3.10 CDL: Zero
Set the zero dead load cancellation. This is analogous with pressing the Zero key on the front of the instrument.
General
No. of parameters 0
Save changes At input
Increment Trade Counter No
Example: Set zero dead load of unit 1
Command Answer Description
S01; Select unit 1
CDL; 0 CRLF Zero dead load set successfully
<load disturbed >
CDL; 1 CRLF Scale in motion
Table of possible replies:
? CRLF Command not understood
0 CRLF Command correctly operated
1 CRLF Scale in motion
2 CRLF Range out. For example, zero setting out of range
3 CRLF System error
4 CRLF Device not ready. For example, printer not ready
21.3.11 CFM: Clock formats
Clock time and date formats.
General
No. of parameters 2
Save changes With TDD1
Increment Trade Counter No
Parameter Details
127 C500-600-2.0.3