EasyManua.ls Logo

Roche cobas b 221 system

Roche cobas b 221 system
368 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Roche Diagnostics May 2009
B-98 Service Manual · Version 9.0
4 Components cobas b 221 system
Touch screen/PC unit
o Enter the serial number of the cobas b 221 system – it is located on the rear panel
of the cobas b 221 system.
e
For more information, see Serial number on page B-6.
o Perform the QC material setup: [Setup] > [QC material].
For cobas b 221<2/4/6> systems
o Check the software version ([Info] > [Software versions]) and perform a software
update, if necessary (h [System] > [Utilities] > [Communication] > [SW
upgrade])
e
For more information, refer to:
Software update on page B-104
cobas b 221 system Reference Manual, chapter "Software modes", section "SW upgrade".
o Check date and time.
o Assign the instrument type and load the default setup.
e
For more information, see Instrument type on page B-6.
O In the menu [Setup] > [Protected Setup] > [Instrument type], select the
instrument type.
O Press the [Load default setup] key to load the default settings of the instrument
stored on the hard disk.
O Press the [Save and exit] key.
o Import the previously exported Setup data: [Setup] > [Load / save setup] >
[Import].
o Print the Status report:
[Info] > [Miscellaneous reports] > [Status report] > [Print status report],
compare it to the previously printed report and adjust settings if necessary.
o Check the following settings and adjust if necessary:
O [Setup] > [Interfaces]: Network, COM 1 and ASTM.
O [Setup] > [Security]: Instrument functions, User management and Profile
management.
O [Setup] > [Parameters] > [Misc. Settings]: Activated for calibrations,
Activated for measurements, QC lock, QC warning, Multirules.
O [Setup] > [Display & Reports] > [Calibration] > [Cal. report]: Sensor report,
Parameter report.
O [Setup] > [Protected setup] > [Miscellaneous settings] > [Cal. intervals U.S.].
o Restore the backup data from the floppy disk:
[Data manager] > [Utilities] > [Backup / Restore] > [Copy from disk]
e
For more information, refer to cobas b 221 system Reference Manual, Section
"Utilities".
o Restore the databases:
[Data manager] > [Utilities] > [Backup / Restore] > [Restore archive]
e
For more information, refer to cobas b 221 system Reference Manual, Section
"Utilities".
o Reinsert the wastewater bottle, the S1 Rinse Solution bottle, the S2 Fluid Pack and
the S3 Fluid Pack (cobas b 221<6> system only).
o Insert a new MSS cartridge (cobas b 221<6> system only).
o Reread the barcodes of the electrodes or insert new electrodes.

Table of Contents

Related product manuals