EasyManua.ls Logo

Roche cobas b 221 system

Roche cobas b 221 system
368 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Roche Diagnostics May 2009
Service Manual · Version 9.0 B-123
cobas b 221 system 5 AutoQC module
Replacing the AQC magnetic valve
Use the reverse order for assembly.
Replacing the AQC magnetic valve
1
Switch off the cobas b 221 system.
2
Remove the back panel of the instrument.
e
For instructions, see Removing the rear panel on page B-20.
3
Pull the AutoQC module out to the left.
4
Remove the cap from the magnetic valve (see below).
5
Unscrew the two screws holding the magnetic valve (see above).
6
Pull the connector cable from the board and remove the magnetic valve.
Use the reverse order for assembly.
Figure B-93 Snap lock (b)
Activate the function program
h
[System] > [Component test] > [Aggregates] > [AQC motors]
> [AQC motors] > [Service Position] to bring the motors to the service position.
A Magnetic valve
B Two screws holding the magnetic valve
C Cap
Figure B-94 AutoQC magnetic valve
A
B
C

Table of Contents

Related product manuals