EasyManua.ls Logo

Roche cobas b 221 system

Roche cobas b 221 system
368 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Roche Diagnostics May 2009
B-10 Service Manual · Version 9.0
3 Protected software functions cobas b 221 system
Protected setup
Polarization voltages
Upol O
2
The keyboard symbol under Upol O
2
can be used to change the polarization voltage of
the oxygen electrode (default value 750 mV). The [Adjust] key is used to confirm and
import the newly defined value.
Upol MSS (cobas b 221<5/6> systems only)
The keyboard symbol under Upol MSS can be used to change the polarization voltage
of the MSS sensors (default value -350 mV). The [Adjust] key is used to confirm and
import the newly defined value. The polarization voltage can only be changed with
the MSS cassette inserted.
COOX module (cobas b 221<2/4/6> systems only)
Setup
The number of samples during a measurement is shown under Scans
(1, 2, 4, 8 and 16 samples are possible).
The duration of sampling during a measurement is shown under Integration time
(12 ms to 12000 ms are possible).
COOX switches
The [Suppress negative values] key activates (green) or deactivates (red) the
suppression of the output of negative values during a measurement. The
measurement values are set to 0 in the event of activation and a matching tHb
compensation is made.
COOX adjustments
The offset values for the polychromator bilirubin are stored here. After replacing the
polychromator bilirubin, the touch screen/PC unit or the hard disk, they must be re-
entered by means of entering the code. The code is located on the polychromator and
must be entered using the keyboard symbol under Set.
The [Set default value] key loads the default settings of the polychromator bilirubin
stored on the harddisk.
AutoQC module
The [Temperature correction] key activates (green) or deactivates (red) the
temperature correction of the pH, PO
2
and PCO
2
measurement values during a QC
measurement to receive standardized measurement results.
This setting is intended for internal purposes only and must be changed only when specifically
directed!

Table of Contents

Related product manuals