Roche Diagnostics May 2009
Service Manual · Version 9.0 B-9
cobas b 221 system 3 Protected software functions
Protected setup
Default maintenance schedule:
Unmodified settings:
o User management
o Profile management
o Measurement reports
o Instrument type
o Network settings
o QC material setup and QC timing
o Sample counter
o COOX adjustments
o Serial number
o AQC module
o AQC temperature correction
o Bili
o Units
AutoQC mat assignment all settings deleted
Service mode OFF
PC Demo mode OFF
Maintenance Remind. Interval Samples
BG cleaning No Never 500
ISE cleaning No Never 200
MSS cleaning No Never ---
COOX cleaning No Never 20
tHb cleaning No Never 20
Change SD cartridge No 1095 days ---
Change MC cartridge No 1095 days ---
Change Hb cartridge No 1095 days ---
Change main pump tubing No 365 days ---
Change MSS output pump
tubing
No 365 days ---
Change MSS input pump
tubing
No 365 days ---
Change FMS tubing No 365 days ---
Change V6, V13, V18 tubing No 365 days ---
Change fill port No 365 days ---
Change T&D tubing No 365 days ---
Change AQC steel tube No 365 days ---
Change AQC barex tube No 365 days ---
Change AQC wash port No 365 days ---
Ta bl e B -3 Default maintenance schedule
Setting Default Comment
Ta bl e B -2 Default settings