Roche Diagnostics May 2009
B-8 Service Manual · Version 9.0
3 Protected software functions cobas b 221 system
Protected setup
Economy mode: weekly: start day Friday
Economy mode: weekly: start time 18:00
Economy mode: weekly: stop day Moday
Economy mode: weekly: stop time 05:00
Aspirate from syringe ON
Auto patient ID OFF
Auto save results ON
Apply corrections ON
Supress negative values ON
Display – message box 30 sec
Switch to measurement report 10 sec
Display – measurement report 10 sec
Automatic menu back 3600 sec
Remove sample cont. warning 1 10 sec
Remove sample cont. warning 2 20 sec
Log off user 900 sec
Sound ON
Vo lu m e 3
Cal. intervals U.S. OFF
SW upgrade: Server 0.0.0.0
SW upgrade: Path -----
SW upgrade: User anonymous
SW upgrade: Password 0
Language German
Date format TT.M.JJJJ
Time format HH:MM
Log monitor prinout OFF
Log file generations 5
Log file entries 30000
Module priority BG-COOX-ISE-MSS
Activated for calibrations ON for all measurement
parameters
Activated for measurements ON for all measurement
parameters
QC lock ON for all measurement
parameters
QC warning OFF for all measurement
parameters
activated Multirules rule 1 and rule 2 for all measurement
parameters
AutoQC timing all times deleted
Setting Default Comment
Ta bl e B -2 Default settings

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche cobas b 221 system and is the answer not in the manual?

Roche cobas b 221 system Specifications

General IconGeneral
BrandRoche
Modelcobas b 221 system
CategoryMedical Equipment
LanguageEnglish

Related product manuals