EasyManua.ls Logo

Roche cobas b 221 system

Roche cobas b 221 system
368 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Roche Diagnostics May 2009
Service Manual · Version 9.0 D-101
cobas b 221 system 7 Troubleshooting
Important test routines
o The hemolyzer is damaged: replace unit.
I
SP
: Current with sample and power
Status not ok – :
o The hemolyzer is damaged: replace unit.
o The HSIN board is defective: replace board.
U
SP
: Voltage with sample and power
Status not ok – :
o The hemolyzer is damaged: replace unit.
o The HSIN board is defective: replace board.
References
e
Replacing the hemolyzer on page B-60.
e
Replacing the HSIN board on page B-61.
Control sensors
Vacuum System
h At the ready screen, press
[System] > [Component test] > [Control sensors] > [Vacuum system].
There are four effective test options for the vacuum system that can be used for error
analysis:
o Set 200 mbar as target value and switch on pump: the underpressure should be
between 170 and 230 mbar after 2 to 3 seconds. If this cannot be reached, valve
V13 or the connecting tube must be checked and if necessary replaced.
o Set 400 mbar as target value and switch on pump: the underpressure should be
between 360 and 440 mbar after 10 seconds. If this cannot be reached, valve V13
or the connecting tube must be checked and if necessary replaced.
o Set "max" as target value and switch on pump: the underpressure should reach at
least 75% of the local Barometric Pressure as sensed on Control sensors/Baro
sensor. If this is not reached, check the pump head and the VPS sensor and if
necessary replace.
o Set "max" as target value, switch on pump and then off after 15 seconds: the
underpressure drop should be less than 50 mbar after 30 seconds. If this is not
reached, there is a leak in the vacuum system: check tightness of vacuum pump,
tube connections, moisture trap, VPP, V13, V18, V6, V23, main pump, waste
separator and if necessary replace.
Baro sensor
h At the ready screen, press
[System] > [Component test] > [Control sensors] > [Baro sensor].
The value of the Baro sensor must match the actually measured Baro value (with a
calibrated barometer). If the deviation is greater than 4 mbar, the Baro value must be
input again in the protected setup.
e
For instructions, see Baro settings on page B-5.

Table of Contents

Related product manuals