Roche Diagnostics May 2009
B-12 Service Manual · Version 9.0
3 Protected software functions cobas b 221 system
Protected infos
3
Press h [Activate system data].
The instrument software will be restarted and the restored system data is available.
AQC adjustment
The start [Start] key initializes a position compensation of the AutoQC motors to
ensure that ampoules are pierced at the correct position after replacement of the
AutoQC board. This should be performed in the course of AQC maintenance.
Protected infos
Service log file export
This function exports log files to USB mass storage or to a floppy disk.
The naming of the logfile archive is as follows:
logfile_xxxx_yyyyyy_zzzz.pck
x...serial number
y...date
z...time
In case of USB, the destination folder is Roche\log\.
To avoid damage, the AQC ampoule mats must be removed from the AQC ampoule block before
starting the adjustment!
Be aware of possible injuries due to moving parts!
USB functionality
When USB mass storage devices (USB flash drives) are connected to the instrument export and
import will use USB instead of floppy disk. Export to and import from USB mass storage requires
software version 5.00 or higher to be installed on instruments using touch screen/PC units version
4.X or higher and interface unit cobas b 221 system > SN 3000.
When there is too little free space on the floppy disk, data can be deleted. If this is confirmed with
„Yes“, all existing data on the floppy disk will be deleted.
The log file may not be exported during any active procedures (calibration, measurement, fill
routines, diagnostics, ...).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche cobas b 221 system and is the answer not in the manual?

Roche cobas b 221 system Specifications

General IconGeneral
BrandRoche
Modelcobas b 221 system
CategoryMedical Equipment
LanguageEnglish

Related product manuals