EasyManua.ls Logo

Roche cobas b 221 system

Roche cobas b 221 system
368 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Roche Diagnostics May 2009
B-122 Service Manual · Version 9.0
5 AutoQC module cobas b 221 system
Replacing the AQC snap lock
7
Pull the AutoQC module out to the left and remove it.
Use the reverse order for assembly.
Replacing the AQC snap lock
The snap lock of the cobas b 221 system AutoQC module consists of two parts. Part a
is fixed to the AutoQC module; part b is fixed to the right frame of the cobas b 221
system.
a To replace the AQC snap lock (a)
1
Remove the back panel of the instrument .
e
For instructions, see Removing the rear panel on page B-20.
2
Pull the AutoQC module out to the left.
3
Unscrew the two screws holding the snap lock on the back of the AutoQC module
and remove them (see below, A).
Use the reverse order for assembly.
a To replace the snap lock (b)
1
Remove the back panel of the instrument.
e
For instructions, see Removing the rear panel on page B-20.
2
Unscrew the four screws holding the right side panel and remove the side panel.
3
Pull the AutoQC module out to the left.
4
Press the brackets on the back of the snap lock and slide it through the opening
towards the AutoQC module.
A Two screws holding the snap lock
Figure B-92 Snap lock (a)
A
Do not remove the fastening screws completely, because they are secured against loss.

Table of Contents

Related product manuals