EasyManua.ls Logo

Roche cobas b 221 system

Roche cobas b 221 system
368 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Roche Diagnostics May 2009
A-42 Service Manual · Version 9.0
2 Fluid actions cobas b 221 system
Calibration
CAL 2/CAL 3/CAL 4 Calibration
CAL 2 Calibration This procedure determines the linearity of the sensors. CAL 2 is aspirated from the
S3 Fluid Pack A via the T&D disk (position 6), sample inlet path and sample
distributor into the MSS measuring chamber and measured.
CAL 3 Calibration This procedure determines the interference sensitivity of the sensors. CAL 3 is
aspirated from the S3 Fluid Pack A via the T&D disk (position 8), sample inlet path
and sample distributor into the MSS measuring chamber and measured.
CAL 4 Calibration This procedure determines the interference sensitivity of the sensors. CAL 4 is
aspirated from the S3 Fluid Pack A via the T&D disk (position 11), sample inlet path
and sample distributor into the MSS measuring chamber and measured.
A MSS output pump
B T&D module (CAL 2=Pos.6, CAL 3=Pos.8, CAL 4= Pos.11)
C Waste separator
D CAL 2
E CAL 3
F CAL 4
Figure A-25 CAL 2/CAL 3/CAL 4 Calibration
A
C
B
D E F

Table of Contents

Related product manuals